Specifications

 

Parameter Characteristic Specification
pH / mV Range 0.00 to 14.00 pH* / ±600.0 mV
Resolution 0.01 pH / 0.1 mV
Accuracy ±0.02 pH / ±1.2 mV
Calibration 1 point with HI9828-25 Quick Calibration solution. Up to 3 points with standard or customizable values.
ORP Range ±2000.0 mV
Resolution 0.1 mV
Accuracy ±1.0 mV
Calibration Automatic to a customizable value (relative mV)
Dissolved Oxygen Range 0.0 to 500.0 % saturation; 0.00 to 50.00 ppm (mg/L)
Resolution 0.1 % saturation; 0.01 ppm (mg/L)
Accuracy ±1.5 % (from 0.0 to 200.0 % saturation), ±5 % (over 200.0 % saturation)
Calibration 1 point with quick calibration in saturated air or customizable solutions.
Pressure Compensation Automatic (450 to 850 mmHg)
EC (Conductivity) Range 0 to 200 mS/cm; 0 to 400 mS/cm (absolute)
Resolution Automatic or manual depending on the scale
Accuracy ±1 % of the reading or ±1 μS/cm, whichever is greater
Calibration Standard or customizable solutions with HI9828-20
Resistivity Range 0 to 999999 Ω·cm; up to 1.0000 MΩ·cm
Resolution 1 Ω·cm; 0.1 kΩ·cm; 0.0001 MΩ·cm
Calibration Based on conductivity calibration
TDS Range 0 to 400,000 ppm (mg/L)
Resolution Manual or automatic depending on the scale
Accuracy ±1 % of the reading or ±1 ppm (mg/L), whichever is greater
Calibration Based on conductivity or salinity calibration
Salinity Range 0.00 to 70.00 PSU
Resolution 0.01 PSU
Accuracy ±2 % of the reading or ±0.01 PSU, whichever is greater
Calibration 1 point, using a customizable solution
Seawater Sigma Range 0.0 to 50.0 σt, σ₀, σ₁₅
Resolution 0.1 σt, σ₀, σ₁₅
Accuracy ±1.0 σt, σ₀, σ₁₅
Calibration Based on conductivity or salinity calibration
Turbidity Range 0.0 to 99.9 FNU; 100 to 1000 FNU
Resolution 0.1 FNU (up to 99.9 FNU); 1 FNU (over 100 FNU)
Accuracy ±0.3 FNU or ±2 % of the reading, whichever is greater
Calibration Automatic, up to three points
Atmospheric Pressure Range 450.0 to 850.0 mmHg; other units included (inHg, mbar, psi, atm, kPa)
Resolution 0.1 mmHg; other resolutions based on the unit
Accuracy ±3.0 mmHg within ±15 °C from temperature calibration
Calibration Automatic to a customizable point
Temperature Range −5.00 to 50.00 °C; 23.00 to 122.00 °F; 268.15 to 323.15 K
Resolution 0.01 °C; 0.01 °F; 0.01 K
Accuracy ±0.15 °C; ±0.27 °F; ±0.15 K
Calibration Automatic to a customizable point